| Model Number |
SUTR ANCH, BIO-COMP P - LCK, 2.4 X 11.3 |
| Medical Device Problem Codes |
Break (1069); Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a labrum refixation with biceps surgery the implant did not hold in the patient and got frayed at the end.All fragments were retrieved from the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-2922bc-1 suture anchor, biocomposite¿ pushlock® 2.4 x 11.3 mm serial/batch number (b)(6) was received for evaluation.Functional testing moving the inner shaft inside the outer shaft found that the devices smoothly engage.No damaged or issues were found.A visual inspection revealed that the bio screw was broken off.A piece of screw on the shaft shows bents on the threads.The method of bone preparation employed during the procedure and the bone quality encountered were not provided.The most likely causes of the reported failure are improper bone preparation, misaligned insertion, and prying/leveraging the device during insertion.Direction for use.Dfu-0087.G.Precautions.2.Surgeons are advised to review the product-specific surgical technique prior to performing any surgery.Arthrex provides detailed surgical techniques in print, video, and electronic formats.The arthrex website also provides detailed surgical technique information and demonstrations.Or contact your arthrex representative for an onsite demonstration.3.Make sure to use the recommended drill bit or punch to create the bone socket.4.Pushlock and swivelock suture anchors only: during anchor insertion, ensure that the angle of anchor insertion is coaxial to that of the previously prepared bone socket.5.Pushlock and swivelock suture anchors only: insert the driver into the bone socket until the anchor body makes contact with the bone.Preview and adjust suture tension, if necessary.Tension will not increase during final advancement of the anchor body.6.Pushlock and swivelock suture anchors only: ensure that the anchor body is in full contact with the bone before advancing the anchor body into the prepared bone socket.8.Self-punching pushlock and swivelock suture anchors only: ensure that the angle of anchor insertion is perpendicular to the bone.
|
| |
|
Search Alerts/Recalls
|