| Catalog Number |
6720240007 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The device tip was found to be fractured off and missing from the device, posing the risk of a small component being lost in a surgical site, during a service evaluation at the manufacturer facility.There were no adverse consequences related to this event.
|
| |
|
Event or Problem Description
|
|
The device tip was found to be fractured off and missing from the device, posing the risk of a small component being lost in a surgical site, during a service evaluation at the manufacturer facility.There were no adverse consequences related to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: follow-up report submitted to document device evaluation results.
|
| |
|
Search Alerts/Recalls
|