• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LATITUDE PROGRAMMING SYSTEM; EXTERNAL PROGRAMMER Back to Search Results
Model Number 3300
Medical Device Problem Codes Defective Device (2588); Application Program Problem (2880)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This programmer was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Visual inspection of the device identified no abnormalities.The programmer was powered on, and it was confirmed the touchscreen was intermittently unresponsive due to not detecting touch inputs.No error or fault codes were recorded in the programmer's logfile.The programmer was disassembled, and it was determined that the cause of the touchscreen becoming unresponsive during use was due to the water detection feature within the programmer's lcd touchscreen assembly.Various inputs can induce false positive detections of this water detection feature, which then results in an unresponsive touchscreen.An unresponsive touchscreen is a known and anticipated potential hazard that can occur during use of the device which is accounted for in device risk documentation and labeling.Boston scientific has received similar reported events where the latitude programmer screen became unresponsive to touch, and a complaint review confirmed that the occurrence of harm associated with these events is low.Investigation of this behavior has not found any commonalities (i.E., device serial numbers, date of the event, date of manufacture, etc.) in the reports received.
 
Event or Problem Description
It was reported that this programmer displayed an error code.No patient involvement was reported.The programmer was returned to boston scientific for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE PROGRAMMING SYSTEM
Common Device Name
EXTERNAL PROGRAMMER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
PLEXUS ELECTRONICA S DE RL DE CV
paseo del norte
no 4640 guadalajara
zapopan, jal 45010
MX   45010
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key20754021
Report Number2124215-2024-73602
Device Sequence Number13084389
Product Code OSR
UDI-Device Identifier00802526616105
UDI-Public00802526616105
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P910077/S159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 11/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3300
Device Catalogue Number3300
Device Lot Number009081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2024
Initial Report FDA Received Date11/21/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-