| Model Number |
GF-UCT180 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was observed during the device inspection, there were foreign objects on the forceps elevator wire of the gastrovideoscope.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.Additional information: h3, h4, h6, h11 a review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, olympus confirmed foreign material on the wire, but the type of the material or root cause cannot be identified the event can be detected/prevented by the following instructions for use which state: chapter 6 compatible reprocessing methods and chemical agents chapter 7 cleaning, disinfection, and sterilization procedures olympus will continue to monitor field performance for this device.
|
| |
|