ETHICON INC. SFXSPI PDS+ BI VIO 9IN2 2-0 DA CT-1; STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE Back to Search Results
Catalog Number SXPP2B408
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information: h6 component code: g07002 - device not received.Additional information provided: customer will return the 6 needles along with the 4 remaining sutures within the same lot.Same thing had happened to 3 of them.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported a patient underwent an unknown procedure on an unknown date in 2024 and suture was used.During a case that when they were attempting to tighten the suture it would pop off from the needle.Surgeon opted to use a basic suture to complete the case without any patient harm.No needles fell into the patient.Additional information has been requested.
 
Additional Manufacturer Narrative
(b)(4) this report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9 h3, h6 additional h6 component code: g07002 incomplete device returned additional h6 component code: g07002 no device problem found.Investigation summary: the product was returned to ethicon for evaluation.Two detached needles were received for analysis.The product code sxpp2b408 which is a strand double armed.The visual assessment of the needles noted that the swage and attachment area were observed as expected.However, marks were noted on the body.The barrel hole of the needle was examined under magnification, and no suture remnant was found.The suture was not returned for evaluation to determine the assignable cause.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional investigation summary of device from same lot returned from user: the product was returned to ethicon for evaluation.One open box with one representative unopened sample was received for analysis.Product code sxpp2b408.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packet.In order to evaluate the conditions of the returned samples, the packet was opened, and no defects were detected.The swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment and the pull force result was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFXSPI PDS+ BI VIO 9IN2 2-0 DA CT-1
Common Device Name
STRATAFIX SPIRAL PDS PLUS BIDIRECTIONAL KNOTLESS TISSUE CONTROL DEVICE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo 00754
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key20759319
Report Number2210968-2024-12362
Device Sequence Number20972583
Product Code NEW
UDI-Device Identifier10705031464810
UDI-Public(01)10705031464810(17)250430(10)TEBBKB
Combination Product (Y/N)Y
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K192144
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Catalogue NumberSXPP2B408
Device Lot NumberTEBBKB
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2024
Supplement Date Received by Manufacturer11/30/2024
Initial Report FDA Received Date11/22/2024
Supplement Report FDA Received Date12/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-