| Model Number |
G22613 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.A possible contributing factor to premature band deployment includes allowing the trigger cord to become lodged between the barrel and the distal end of the endoscope.This can restrict trigger cord movement and result in premature band deployment if enough tension is applied to pull the cord free.Premature band deployment can also occur if the handle is placed in the firing position before the endoscope is in place inside the patient or even while winding the trigger cord.The user is advised not to apply tension to the trigger cord while winding it on the handle to avoid premature deployment of bands.The instructions for use caution the user: ¿with handle in two-way position, slowly rotate handle clockwise to wind trigger cord onto handle spool until it is taut.Note: care must be taken to avoid deploying a band while winding trigger cord." prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
| |
|
Event or Problem Description
|
|
During a varices ligation, the physician used a cook 4 shooter saeed multi-band ligator.It was reported that during the procedure the user found two more varices to be ligated so they introduced the endoscope with barrel on.Two (2) bands were deployed without even turning the handle which wasn't in the firing position.They then managed to ligate with the last two bands.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Search Alerts/Recalls
|