No device was returned for evaluation and no photographs were provided for review.Further, operative notes were not provided for review of usage/technique.A review of manufacturing records was unable to be performed as the lot information of the product involved in the event was not available.Based on the information obtained the complaint was unable to be confirmed and a definitive root cause was unable to be determined; however, the event may have been the result of excessive force being used while securing the knob component, leading to component damage.Labeling review: ".Pre-operative warnings: do not use these instruments for any action for which it was not intended such as hammering, prying, or lifting.The instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury.Instruments should be protected during storage and from corrosive environments.Refer to cleaning and sterilization instructions below for all non-sterile parts.Care should be used during surgical procedures to prevent damage to the device(s) and injury to the patient." ".Intra-operative warnings: the physician should take precautions against putting undue stress on the spinal area with instruments.Any surgical technique should be carefully followed.It is important that the surgeon exercise extreme caution when working in close proximity to vital organs, nerves, or vessels, and that the force applied to the instrumentation is not excessive, to prevent potential injury to the patient.Over-bending, notching, striking, and/or scratching of implants with any instrument should be avoided to reduce the risk of breakage.".
|