The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the mildly tortuous and moderately calcified anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that the procedure was to treat an 88% stenosed lesion in the left circumflex (lcx) artery with moderate calcification and mild tortuosity.The 3.5x12mm nc trek balloon dilatation catheter (bdc) was advanced to the target lesion with resistance due to anatomy.The balloon was inflated; however, the balloon ruptured during the first inflation at 12 atmospheres (atm).Therefore, the bdc was removed from the patient.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|