| Model Number |
VTICM5_12.6 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a refractive surprise.In a separate visit the lens was exchanged for the same size different power.The problem was resolved.Cause reported as unknown.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
Over/under correction and secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h3-device evaluation: lens returned in a microcentrifuge vial in liquid with residue/debris on product.Visual inspection found no visible damage to the lens.Dimensional inspection found the lens to be within specifications.Functional inspection found the lens to not be within specifications.H6-investigation type 4110: lens work order search- no similar complaint event(s) within associated lots were found.H6- method code 3331: device history record (dhr) review: based on the results of the investigation, all released devices from the associated work orders(s), including the suspected device, have been manufactured within established parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|