• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. POLYAXIAL SCREW, DIA. 6.5MM X 40MM; VITALITY SPINAL FIXATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. POLYAXIAL SCREW, DIA. 6.5MM X 40MM; VITALITY SPINAL FIXATION SYSTEM Back to Search Results
Catalog Number 701M6540
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a vitality closure top migrated out of it's mating polyaxial screw post-operatively.The original construct was an l3/4 plif using a competitive cage.One month postoperatively, it was confirmed that the l3 left set screw was loosening, and three months after the surgery, it was confirmed that the set screw was completely popped off.A revision surgery was performed.This is report one of two for this event.
 
Additional Manufacturer Narrative
D4: the remainder of the udi number is unknown because the lot number is not available.Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Additional Manufacturer Narrative
If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.Reference report 3012447612-2024-00369.
 
Event or Problem Description
It was reported that a vitality closure top migrated out of it's mating polyaxial screw post-operatively.The original construct was an l3/4 plif using a competitive cage.One month postoperatively, it was confirmed that the l3 left set screw was loosening, and three months after the surgery, it was confirmed that the set screw was completely popped off.A revision surgery was performed.This is report one of two for this event.
 
Additional Manufacturer Narrative
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.H6: additional method code: 4110.Corrections in h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed no signs of damage.X-ray images were provided which show that a closure top has migrated out of its screw.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to not fully seating the closure tops against the rods.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Event or Problem Description
It was reported that a vitality closure top migrated out of it's mating polyaxial screw post-operatively.The original construct was an l3/4 plif using a competitive cage.One month postoperatively, it was confirmed that the l3 left set screw was loosening, and three months after the surgery, it was confirmed that the set screw was completely popped off.A revision surgery was performed.This is report one of two for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYAXIAL SCREW, DIA. 6.5MM X 40MM
Common Device Name
VITALITY SPINAL FIXATION SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key20765982
Report Number3012447612-2024-00368
Device Sequence Number15778594
Product Code NKB
UDI-Device Identifier00880304867505
UDI-Public(01)00880304867505(10)T11583
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K171907
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number701M6540
Device Lot NumberT11583
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date11/22/2024
Supplement Report FDA Received Date01/29/2025
03/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-