|
Reportable based on device analysis completed on 05nov2024.It was reported that balloon damaged occurred.The 85% stenosed target lesion was located in the non-tortuous and severely calcified artificial arteriovenous fistula in the left upper limb.An 8.0 x40, 75cm gladiator elite balloon catheter was advanced for dilation.The balloon was inflated at 18atm once; however, the pressure was less than bursting pressure, and the pressure was found to be reduced.The balloon was pulled out, and it was found out that the balloon was damaged.The procedure was completed with another of the same device.No complications were reported, and the patient was stable post procedure.However, device analysis revealed a balloon longitudinal tear.
|