| Model Number |
H74939171051510 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b - additional pro code (product code): lit.G4 - additional premarket / 510(k): k110122, k141521.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the superficial femoral artery.A 5.0 x 150, 135cm mustang balloon catheter was advanced for dilation.During inflation, balloon ruptured when the bursting pressure was not reached.The procedure was completed with another of the same device.No complications were reported, and the patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
D2b - additional pro code (product code): lit g4 - additional premarket / 510(k): k110122, k141521.The device was returned for evaluation.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A balloon longitudinal tear was identified beginning at the proximal markerband and extending approximately 68mm distally across the balloon material.The rated burst pressure for this device as per specification is 24 atmospheres.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual and tactile examination found no damage to the tip of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged in the correct position on the device.No other issues were identified with the device and no patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the superficial femoral artery.A 5.0 x 150, 135cm mustang balloon catheter was advanced for dilation.During inflation, balloon ruptured when the bursting pressure was not reached.The procedure was completed with another of the same device.No complications were reported, and the patient was stable post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
B5.Describe event or problem: added information, based on additional information d2b - additional pro code (product code): lit, g4 - additional premarket / 510(k): k110122, k141521, the device was returned for evaluation.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A balloon longitudinal tear was identified beginning at the proximal markerband and extending approximately 68mm distally across the balloon material.The rated burst pressure for this device as per specification is 24 atmospheres.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual and tactile examination found no damage to the tip of the device.A visual examination observed no issues with the markerbands of the device.Markerbands were undamaged in the correct position on the device.No other issues were identified with the device and no patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in the superficial femoral artery.A 5.0 x 150, 135cm mustang balloon catheter was advanced for dilation.During inflation, balloon ruptured when the bursting pressure was not reached.The procedure was completed with another of the same device.No complications were reported, and the patient was stable post procedure.It was further reported that the target lesion was 80% stenosed, non-tortuous and severely calcified.
|
| |
|