• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA PARTIAL TIBIAL CEMENTED SIZE G RIGHT MEDIAL; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. PERSONA PARTIAL TIBIAL CEMENTED SIZE G RIGHT MEDIAL; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/23/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 42558000502 - partial femur cemented size 5 right medial - 64944836.42528200708 - partial articular surface right medial size g 8 mm thickness - 64660649.G2: foreign country: new zealand.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: overall sizing of the implant is appropriate.Possible polyethylene wear of the medial compartment.Early loosening of the femoral and tibial components not excluded.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient underwent a knee arthroplasty.Subsequently 2.5 years post-op, the patient was revised due to pain.Attempts have been made and all available information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA PARTIAL TIBIAL CEMENTED SIZE G RIGHT MEDIAL
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key20768414
Report Number3007963827-2024-00470
Device Sequence Number15600044
Product Code HSX
UDI-Device Identifier00880304812734
UDI-Public(01)00880304812734(17)310523(10)65032727
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K161592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42538000702
Device Lot Number65032727
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2024
Initial Report FDA Received Date11/24/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexPrefer Not To Disclose
-
-