| Catalog Number |
364378 |
| Medical Device Problem Code |
Volume Accuracy Problem (1675)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1.Initial reporter facility name: (b)(6).Investigation summary: "material #: 364378, lot/batch #: 3227183.Bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to missing scale markings as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode missing scale markings.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.".
|
| |
|
Event or Problem Description
|
|
It was reported prior to using a bd a-line¿ syringe, it was observed the syringe barrel did not have scale markings.There was no report of adverse impact to patients or users.
|
| |
|
Event or Problem Description
|
|
It was reported prior to using a bd a-line¿ syringe, it was observed the syringe barrel did not have scale markings.There was no report of adverse impact to patients or users.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental mdr is submitted to document that bd will no longer consider the specific malfunction identified below to be a reportable mdr malfunction for the specified product family.Consistent with fda's "medical device reporting for manufacturers" guidance (november 8, 2016, section 2.15), bd determined that this issue has not caused or contributed to a death or serious injury in the previous two years and that the likelihood of a death or serious injury occurring as a result is remote.Bd previously notified fda of this determination on january 8, 2026.Product family: bd a-line¿
specific malfunction: smeared / smudged scale marking on the device where volumetrics are not readable
this mdr is being submitted as a representative notification for the product family and specific malfunction identified above and applies to all applicable catalog numbers within that product family associated with this issue.This determination applies only to the issue identified above when it occurs independently and does not otherwise meet mdr reporting requirements.Bd will continue to submit mdrs for reportable deaths, serious injuries, and reportable malfunctions, including events involving multiple event types where mdr reporting criteria are met.Future occurrences of this issue will no longer be submitted as standalone malfunction mdrs.Bd will continue to monitor complaint and adverse event data related to this issue and will reassess mdr reportability should new information indicate that reporting criteria may be met.
|
| |
|