|
|
| Model Number |
H74939407XS0 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This medwatch report is being filed based on the images and video received on november 18, 2024.The supplied image and video presents a ductile, tensile overload fracture of the core wire, with the distal tip of the safari2 guidewire lodged in a non (b)(6) medical device with subsequent coil stretching.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at (b)(6) medical and was determined to be acceptable.As described in the device instructions for use (dfu) preparation for use: ·inspect and prepare the catheter according to the manufacturer's instructions.This includes flushing the catheter to be used with heparinized saline solution.Verify compatibility of interacting devices prior to use.If coils of a guidewire separate, do not remove the core.Carefully remove the coils and core simultaneously.As described in the device instructions for use (dfu): 1.Ensure a catheter is properly placed in the ventricle or other intended treatment area.2.Carefully remove the safari2tm guidewire from the hoop by grasping the proximal end of the j-straightener separating the straightener from the hoop.3.After inspection of the safari2tm guidewire, advance the j-straightener over the distal section of the safari2tm guidewire to straighten the curve.4.Insert the safari2tm guidewire into the hub of the catheter with the aid of the j-straightener.5.Carefully advance the distal tip of the safari2tm guidewire into the lumen of the catheter.6.Remove the safari2tm guidewire j-straightener by withdrawing it over the length of the safari2tm guidewire.7.Advance the safari2tm guidewire through the catheter under fluoroscopic guidance.As described in the device instructions for use (dfu) warnings: exercise care in handling of the guidewire during a procedure to reduce the possibility if accidental breakage, bending, kinking or coil separation.At this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided and the evidence presented, it appeared that procedural and/or handling factors have contributed to the event as reported.Should additional information be received, a follow up medwatch report will be submitted.Note: although annex g instructions state that with stand-alone devices annex g term should not be used, code 4755 was selected due to component code (g) showing up on the missing data report.
|
| |
|
Event or Problem Description
|
|
Event description: during the access to la, specifically in the transeptal punction the safari wire the coil of the safari wire is damaged, it comes off and opens as if stretching.What troubleshooting steps, if any, resolved the issue? remove the guide and try using another alternative guide.Patient present at time of event? yes.Patient complications: no patient complications.Question: photo or video of issue? answer: yes.Question: is it known what caused the device damage? answer: maybe during the insertion.Question: what were possible contributing factors (comorbidities, anatomy characteristics, etc)? answer: i don´t know.Question: when specifically, during device prep or during the procedure did the device damage occur? answer: during the procedure.Question: where specifically on the device did the damage occur? answer: when the doctor across the interacuricular septum.Question: device/procedure outcome: answer: procedure completed via alternative method.
|
| |
|
Search Alerts/Recalls
|
|
|