| Model Number |
WA33036A |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted, when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported, that the subject device had the lens cracked.The issue was found, during preparation for use.For an unspecified procedure.The procedure was completed using a similar device.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.Updated fields: e1, h4, h6, h11 the device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction; blurry/ dark image caused by cracked lenses, component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information from customer.
|
| |
|