SMITH & NEPHEW, INC. JOURN ART INS BCS STD 3-4 RT 9; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
74023231 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Arthritis (1723); Failure of Implant (1924); Joint Laxity (4526)
|
| Date of Event |
11/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, after a tka in both knees, the patient suffered from bilateral instability and knee pain.This adverse event was solved by revision surgery of the right side on (b)(6) 2024 in which the journ art ins bcs std 3-4 rt was explanted due to broken and worn post.It was also found that the patient has progression of osteoarthritis of the patella.Current health status of patient is unknown.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h6 (type of investigation), h7, h10 (roi).H3, h6: the device was not returned for evaluation; therefore, a device analysis could not be performed.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.A review of previous actions concluded that there is a higher-than-expected risk of revision surgery for the first generation of journey bcs systems.It was recommended to undertake a field safety corrective action to inform surgeons of the higher expected risk of revision, and ensure clear communication that the device should only be used for polyethylene exchange revision of journey bcs total knee constructs where the femoral component and tibial baseplate are well fixed.According with the inspection drawing, a verification that parts are free of or damaged areas, burrs, inclusions, porosities, sharp edges, nicks or cracks shall be performed.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include patient anatomy, abnormal loading of limb, excessive forces and/or injury.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.H6 (investigation findings).
|
| |
|