• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5 13.2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Iritis (1940)
Date of Event 11/06/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Iritis is an inflammatory condition which is common after intra-ocular surgery, however this is usually mild and does not require any additional interventions beyond the standard post-op regimen.In some cases, the degree of inflammation may be more severe and require extended steroid treatment.Procedural factors including excessive surgical manipulation and the use of proinflammatory substances in the eye may have contributed to the event.Mechanical insult to the iris can result in a prolonged or stronger inflammatory response.An exact cause could not be determined as the event could be multifactorial in nature including patient and/or procedure related factors.H6 - investigation type 4110: lens work order search - no similar complaint event(s) within associated lots were found.Claim# 738390.
 
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced low vault and recurrent iritis.Lens was explanted.Cause of event was not reported.
 
Additional Manufacturer Narrative
H3: device evaluation - lens was returned in microcentrifuge tube dry.Visual inspection found no visible damage.Dimensional inspection found the lens to be within specification.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
800 royal oaks dr
suite 210
monrovia, CA 
MDR Report Key20775898
Report Number2023826-2024-05417
Device Sequence Number19936687
Product Code QCB
UDI-Device Identifier00840311311712
UDI-Public(01)00840311311712(11)240310(17)270228
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICM5 13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/06/2024
Supplement Date Received by Manufacturer01/09/2026
Initial Report FDA Received Date11/25/2024
Supplement Report FDA Received Date01/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FOAM TIP PLUNGER, LOT# UNK; MSI-PF, LOT# UNK; SFC-45, LOT# UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-