• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 32 M/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG TRIAL FEMORAL HEAD 32 M/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Catalog Number 75100851
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, during thr surgery, the internal part of one (1) trial femoral head 32 m/+4 broke.It broke into one (1) piece, that was completely removed from the patient and it was verified that no piece remained.The procedure was resumed, without any delay, using the same device.No injury was reported as a consequence of this issue.The patient's health condition is stable.
 
Additional Manufacturer Narrative
Additional information: h6, h10.H3, h6: the device used in treatment was not returned for investigation.A product evaluation was not possible.No batch number was communicated so the production history review was not possible.Due to an unknown batch number, the review of historical complaints was performed on product number basis only, revealing 1 additional complaint over the past 12 months with similar failure mode.Due to insufficient information, it is not possible to perform a review of past corrective actions.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The performed investigation does not lead to an accurately determined cause.Based on the available information it is not possible to investigate whether the reported device met manufacturing specifications upon release for distribution.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from repeated resterilization, prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.Therefore, all reusable instruments must be routinely inspected for signs of wear and damage after reprocessing.Any instruments found to be damaged should be replaced as necessary to ensure their proper functioning and to maintain patient safety.This guidance is provided by smith & nephew orthopaedics ag (reference: lit.N°03389-en 1363 v3 11/19).There is no need for further actions because of the limited information provided.Nevertheless, smith+nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.D4 (lot number).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIAL FEMORAL HEAD 32 M/+4
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key20778269
Report Number9613369-2024-00150
Device Sequence Number12999049
Product Code LZO
UDI-Device Identifier07611996121841
UDI-Public7611996121841
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K132435
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number75100851
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2024
Supplement Date Received by Manufacturer12/04/2024
Initial Report FDA Received Date11/26/2024
Supplement Report FDA Received Date12/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age35 YR
Patient SexFemale
Patient Weight70 KG
-
-