|
Tw # (b)(4).Updated sections: b4, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 11/13/2024.An investigation was conducted on 11/26/2024.A visual inspection was conducted.Both the harvesting device and cannula was returned for evaluation.Signs of clinical use and evidence of blood was observed.There were no visual defects observed on the intact cannula or intact c-ring.There were no visual defects observed on the clear intact silicone insulation on both the cold and hot jaws.The heater wire was observed to be flexed away from the hot jaw from the base of the hot jaw with detachment at the tip of the hot jaw.No other visual defects were observed.Based on the returned condition of the device as well as the evaluation results, the reported failure "material twisted/ bent wire" was confirmed.The lot # 3000422785 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.Specific actions for the reported failure mode are being maintained and documented under maquet's corrective and preventive action (capa) system.
|