| Model Number |
OTV-S7H-NA-10E |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E2/e3: no information available for the occupation of the initial reporter or whether they are a healthcare professional the investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that, the subject device exhibited poor image.The issue was observed during set up before the patient was in room.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for inspection and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: image failure and a pinhole in the cable.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the most probable cause was traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|