| Catalog Number |
VBCR111002A |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/07/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2024, a gore® viabahn® endoprosthesis (vsx device) was intended for use in the cephalic arch during treatment of stenosis.It was reported that the stenosis was pre-dilated with a 4 mm conquest balloon.The physician then advanced the vsx device through a 10fr terumo introducer sheath, over a 0.035" stiff road runner guidewire and across the lesion.Reportedly, after deployment was initiated the deployment line got stuck.The physician attempted to pull the deployment line a few more times with no success and with no device expansion the vsx device was removed.The patient did not experience any adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the manufacturing records indicated the device met pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Additional Manufacturer Narrative
|
|
The viabahn® device was returned to gore for evaluation returned with the endoprosthesis partially expanded.It was reported that this partial expansion occurred when continued deployment was attempted on the back table after withdrawal of the vsx device.The evaluation observed the deployment mechanism to be functional: the deployment line was not caught on any part of the device and deployment continued during evaluation.The cause for the reported deployment difficulty could not be established.Further information regarding this event was requested by gore, but no further information has been reported, therefore this investigation is considered complete.
|
| |
|
Search Alerts/Recalls
|