• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEVRO CORP. SENZA; NEVRO SENZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEVRO CORP. SENZA; NEVRO SENZA Back to Search Results
Model Number LEAD2005-90B
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the patient underwent a revision and it was found there was the white wire broken from the paddle while it was implanted.An hcp assessment regarding the cause of the reported event was not available.The patient had the device explanted.There have been no reports of further complications regarding this event.
 
Additional Manufacturer Narrative
The manufacturing records were reviewed and no relevant non-conformities were found.Nevro is awaiting the return of the device.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Additional Manufacturer Narrative
Updated sections: d9, g1, h2, h3, h6 codes for type of investigation, investigation findings and investigation conclusion.Visual inspection was performed on the returned device and noticeable damage was observed.The microscopy confirmed that some sort of force, strain, or tension likely pulled the white lead out of the paddle.The investigation could not find the cause of the fractured lead as it occurred while implanted in the patient¿s body, and there were no reports of any falls, injuries, surgeries, or other events/activities.It is possible excessive force, strain, or tension pulled the paddle lead apart, which caused the white lead to become disconnected from the paddle.However, a definitive root cause of the fractured lead cannot be conclusively determined with the available evidence.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Common Device Name
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
john paxton
1800 bridge parkway
redwood city, CA 94065
6504332730
MDR Report Key20779880
Report Number3008514029-2024-01082
Device Sequence Number5052220
Product Code LGW
UDI-Device Identifier00813426020640
UDI-Public00813426020640
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Model NumberLEAD2005-90B
Device Lot Number94525321
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2024
Supplement Date Received by Manufacturer05/28/2025
Initial Report FDA Received Date11/26/2024
Supplement Report FDA Received Date06/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-