|
Intuitive surgical, inc.(isi) received the da vinci product to perform failure analysis.The long bipolar grasper instrument was analyzed and found to have damaged conductor wire insulation at the yaw pulley.There was no fragmentation of the insulation, and the internal conductor wires were exposed.Although the conductor wire insulation was damaged, the internal wire was not frayed or fully broken.The instrument failed the electrical continuity test.The complaint regarding damaged conductor wire was confirmed by failure analysis.Blank mdr fields: the missing patient information in sections a and b was either unknown, unavailable, not provided, or not applicable.The expiration date for section d4 is not applicable.Field d6 is blank because the product is not implantable.Information for the blank fields in section e1 is not available.Fields g5 and g7 are not applicable.
|