• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. ACL REPAIR TIGHTROPE, WITH INTERNALBRACE; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ACL REPAIR TIGHTROPE, WITH INTERNALBRACE; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number ACL REPAIR TIGHTROPE, WITH INTERNALBRACE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 11/04/2024, a sales representative reported via phone that an ar-1588btb-ib tightrope ii btb and an ar-1588r-ib acl repair tightrope broke at the button upon tensioning.All broken pieces were removed from the patient, and the case was completed with another ar-1588r-ib acl repair tightrope from the same lot.As a precaution, the surgeon did not tension the tightrope too tightly and put a screw in to hold it securely.There was a 30-minute case delay with additional anesthesia administered to the patient.This was discovered during an acl reconstruction procedure on (b)(6) 2024, with no reported patient harm.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACL REPAIR TIGHTROPE, WITH INTERNALBRACE
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key20781847
Report Number1220246-2024-08608
Device Sequence Number5003955
Product Code MBI
UDI-Device Identifier00888867363656
UDI-Public(01)00888867363656(11)240806(17)290731(10)15298197
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K231857
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACL REPAIR TIGHTROPE, WITH INTERNALBRACE
Device Catalogue NumberAR-1588R-IB
Device Lot Number15298197
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2024
Supplement Date Received by Manufacturer06/12/2025
Initial Report FDA Received Date11/26/2024
Supplement Report FDA Received Date07/08/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-