| Model Number |
ACL REPAIR TIGHTROPE, WITH INTERNALBRACE |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 11/04/2024, a sales representative reported via phone that an ar-1588btb-ib tightrope ii btb and an ar-1588r-ib acl repair tightrope broke at the button upon tensioning.All broken pieces were removed from the patient, and the case was completed with another ar-1588r-ib acl repair tightrope from the same lot.As a precaution, the surgeon did not tension the tightrope too tightly and put a screw in to hold it securely.There was a 30-minute case delay with additional anesthesia administered to the patient.This was discovered during an acl reconstruction procedure on (b)(6) 2024, with no reported patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
|
| |
|