| Catalog Number |
381533 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd insyte autoguard winged is unable to connect.The following information was provided by the initial reporter: the catheter hubs have extra plastic on the luer connector preventing a well seated connection that leads to leaks, requiring the staff to restart ivs.The product has been retained.13nov.It is mentioned "there have been two reported events", was it with different patients? yes.Can you please provide an exact date of event in format dd-mmm-yyyy for both events? 11/7/2024 and 11/5/2024, apparently there were more events before this but these are the 2 that i was made aware of when was this defect observed? during or before use? after; the problem was noticed when the iv tubing was unable to be connected properly.What was the impact to the patient? they had to get stuck again with a new iv any adverse event or serious injury reported to patient or healthcare professional? if yes, please provide the details.Patient was needlessly stuck again.No injury to the staff.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: the complaint of a damaged catheter adapter was confirmed and the cause appeared to be manufacturing related.Two 20g insyte autoguard winged iv catheters were provided for investigation.Deformation was observed in the threaded area of the catheter adapter, which likely prevented a connection with a luer lock device.The appropriate manufacturing personnel were notified of this complaint.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Becton dickinson and company's (bd) medication delivery solutions (mds) business unit will discontinue malfunction mdr reporting for certain device failures that have not caused or contributed to deaths or serious injuries in the past two years, and where the likelihood of a death or serious injury as a result of these malfunctions is remote.This decision follows fda guidelines (medical device reporting for manufacturers guidance for industry and food and drug administration staff issued nov 8, 2016, reference section 2.15).Bd notified fda of this decision on mar 3, 2025.Fda has reviewed and responded to bd¿s notification (reference fda document # (b)(4)) on march 7, 2025.This documentation is available in bd document management system (sap) as (b)(4).This supplemental mdr is being filed to document that the below device failure will no longer be considered a reportable malfunction mdr for the product family below: product family: insyte autoguard.Device failure: molding defective.
|
| |
|