MDT SOFAMOR DANEK PUERTO RICO MFG CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, Back to Search Results
Model Number 6068093
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H6: neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient undergoing posterior f ixation for rod fracture and re-anchoring.Levels implanted at th11-s2ai.It was reported that after final fastening, when attempting to remove the product, it could not be detached from the screw head despite the lock being released. it is believed that the adjacent extender body interfered, leaving no space to move the extender even with the lock released, causing it to get stuck and not detach. a part of the tip was damaged.There was no patient symptom reported.There were no further complications reported regarding the event.Additional information received from manufacturer representative that post-operative infection was confirmed after the initial surgery and cleaning was performed.The date of onset is unknown; no implant malfunctions were confirmed at this time.Afterwards, left rod fracture, loose set screws in the right sacral iliac, and backout of the l5/s cage were confirmed, and repeat surgery was performed on (b)(6) 2024.No patient symptoms are associated with the malfunction.Screw was added for postoperative infections.No defects in the implant itself additional information received from manufacturer representative that there is no causal relationship between post-operative infection and rod, set screw and cage due to implant.Cage backout, rod fracture and set screws did not cause any neurological symptoms.Additional information received from manufacturer representative that there is no malfunction with the extender.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATALYFT¿ PL EXPANDABLE INTERBODY SYSTEM
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT,
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key20783331
Report Number1030489-2024-01568
Device Sequence Number20084999
Product Code MAX
UDI-Device Identifier00763000246723
UDI-Public00763000246723
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K210425
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6068093
Device Catalogue Number6068093
Device Lot Number0944351W
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2024
Initial Report FDA Received Date11/26/2024
Date Device Manufactured01/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-