BARD PERIPHERAL VASCULAR, INC. VACCESS; PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number VA80124
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure in the brachiocephalic vein via superior vena cava junction, the pta balloon allegedly ruptured below the nominal pressure at 4 atm.The procedure was completed using another balloon.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one vacess device was returned for evaluation.An in-house presto inflation device was used to inflate the balloon, and water was noted to be leaking from the balloon.Under microscopic examination, a compound rupture was noted to the balloon.No other functional testing performed.Therefore, the investigation is confirmed for the reported balloon rupture as a compound rupture was noted to the balloon.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: d2b (dqy; lit) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACCESS
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key20783660
Report Number2020394-2024-01743
Device Sequence Number5052328
Product Code LIT
UDI-Device Identifier00801741121265
UDI-Public(01)00801741121265(17)270701(10)93TJ0078
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K111850
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVA80124
Device Lot Number93TJ0078
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2024
Initial Report FDA Received Date11/26/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight35 KG
Patient EthnicityHispanic
-
-