MAUDE Adverse Event Report: UNOMEDICAL A/S MIO; UNO INSET II 60/6 SC1 PINK MIMX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL A/S MIO; UNO INSET II 60/6 SC1 PINK MIMX Back to Search Results
Model Number MMT-923A
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Elevated ketones/Diabetic Ketoacidosis (2364)
Date of Event 08/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
E1: patient city: (b)(6).Patient country: united states.
 
Event or Problem Description
Reference number (b)(4).Event occurred in the united states.It was reported that patient was hospitalized on (b)(6) 2024 due to diabetic ketoacidosis and bent cannula event.Blood glucose at the time of event was 450 mg/dl.The duration of hospitalization was greater than 24 hours.Patient was also found positive for moderate ketones.Patient was treated with insulin via intravenous (iv) drip.Company do not see bent/kinking as being related to human factors, but rather as a training issue including correct choices of insertion sites and infusion sets and cannula length.Furthermore, the soft cannula is a flexible material that during use and upon removal can bend slightly.No further information was available.
 
Additional Manufacturer Narrative
Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary: the information in this complaint (b)(4) has been evaluated.The batch 5404804 in question was manufactured at the reynosa site.Investigation process of the complaint was carried out in accordance with work instruction (wi) guidance for visual testing for complaints area version 3 and wi guidance for functional testing for complaints area version 2 for the code soft cannula found bent upon removal from infusion site complaint investigations.Photo/sample was not provided.In order to test the product, the reference samples from the lot have been requested.Test results: visual test according to wi version 3 on reference samples, reference samples passed the test.Functional (flow) test according to wi version 2 on reference samples, reference samples passed the test.Device history record (dhr) review: the lot 5404804 was manufactured according to the wi version 58, manufactured in the line inset 6, on 12/dec/2022, with a total of (b)(4) units.Review of the dhr showed that all relevant tests required during the related processes had been fulfilled and met the requirements.No deviation were identified; no maintenance events were recorded.Trending: a query was run in database on 18/aug/2025 against malfunction code soft cannula found bent upon removal from infusion site and lot 5404804 and no other complaint have been registered in database for the same lot and malfunction code.Conclusion summary of complaint investigation: as a result of the following: no defect on tests for reference samples related to the complaint, no non-conformance (nc) raised during production related to complaint code, no other complaint received on the lot in question and malfunction code, no further actions are required.This complaint will not require further root cause investigation nor corrective and preventive action (capa) plan.Therefore, this issue will be monitored through the post market surveillance activities.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Additional Manufacturer Narrative
Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary - the information in this complaint (b)(4) has been evaluated.The batch 5404804 in question was manufactured at the reynosa site.Investigation process of the complaint was carried out in accordance with work instruction (wi) guidance for visual testing for complaints area version 3 and wi guidance for functional testing for complaints area version 2 for the code soft cannula found bent upon removal from infusion site complaint investigations.Photo/sample was not provided.In order to test the product, the reference samples from the lot have been requested.Test results: visual test according to wi version 3 on reference samples, reference samples passed the test.Functional (flow) test according to wi version 2 on reference samples, reference samples passed the test.Device history record (dhr) review: the lot 5404804 was manufactured according to the wi version 58, manufactured in the line inset 6, on 12/dec/2022, with a total of (b)(4) units.Review of the device history record (dhr) showed that all relevant tests required during the related processes had been fulfilled and met the requirements.No deviation were identified; no maintenance events were recorded.Trending: a query was run in database on 18/aug/2025 against malfunction code soft cannula found bent upon removal from infusion site and lot 5404804 and no other complaint have been registered in database for the same lot and malfunction code.Conclusion summary of complaint investigation: as a result of the following: no defect on tests for reference samples related to the complaint, no non-conformance (nc) raised during production related to complaint code, no other complaint received on the lot in question and malfunction code, no further actions are required.This complaint will not require further root cause investigation nor corrective and preventive action (capa) plan.Therefore, this issue will be monitored through the post market surveillance activities.Corrective and preventive action (capa) determination results: this complaint falls under the scope of the corrective and preventive action (capa) 1768101: "autosoft 90, kinked soft cannula issues" and will cover inset i (autosoft xc dhf-13.8 & 13.13) and inset ii (autosoft 90 dhf-14 and 14.3) products.Root cause of problem: the root cause has been identified as: method, manpower, measurement, machine: corrective action as a result of the investigation: the action plan and deadlines is as followed: the following actions were already implemented in the non-conformance (nc) 1515780 1.Include the defect in document (b)(4) (work instruction inset line) 2.Improve guards of the conveyors 3.Update trays for the stock of cannulas 4.Update document (b)(4) (quality specification for catheter fixture for skewed catheters) for include the handling of the catheter during the process.5.Update the document (b)(4) (work instruction inset line) to include the preventive actions implemented in the process (trays).A remaining action (to address design root cause) and deadlines is as followed: add cylinders to the lid to maintain in position the needle and cannula during transportation and prior use (database (b)(4) - (b)(6) 2025).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIO
Common Device Name
UNO INSET II 60/6 SC1 PINK MIMX
Manufacturer (Section D)
UNOMEDICAL A/S
aaholmvej 1-3
osted
lejre, 4320
DA  4320
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key20784292
Report Number3003442380-2024-32817
Device Sequence Number8948955
Product Code FPA
UDI-Device Identifier05705244019942
UDI-Public05705244019942
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/10/2025
Device Model NumberMMT-923A
Device Lot Number5404804
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2024
Supplement Date Received by Manufacturer09/03/2025
09/12/2025
Initial Report FDA Received Date11/26/2024
Supplement Report FDA Received Date09/05/2025
10/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-