| Catalog Number |
EVO-10-11-6-B |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Vomiting (2144); Obstruction/Occlusion (2422)
|
| Date of Event |
06/29/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016 date of first implant procedure with evolution® biliary stent system-uncovered evo-10-11-6-b, lot c1237074 in common bile duct.No adverse events related to the procedure.Re-current biliary obstructions 29-jun-2017.Due to tissue or tumor ingrowth.Patient presented with abdominal pain and vomiting.Intervention performed 294 days post procedure.Treatment endoscopic intervention cleaning of prothesis and antibiotics.The site determined the event to not be related to the study device or procedure and noted that relatedness to other circumstances was not documented.(b)(4).Re-current biliary obstructions 23-aug-2018.Due to tissue or tumor ingrowth.Patient presented with abdominal pain.Intervention performed 714 days post procedure.Treatment endoscopic intervention new plastic stent inside the study stent with antibiotics.The site determined the event to be not related to the study stent or procedure, related to stenosis of cancer.(b)(4).Re-current biliary obstructions 14-nov-2019.Due to sludge.Patient presented with abdominal pain.Intervention performed 1162 days post procedure.Treatment endoscopic intervention replacement of plastic stent by study stent inside the previous study stent with antibiotics.The site determined the event to be not related to the study stent or procedure and noted that relatedness to other circumstances was not documented.Patient death (b)(6) 2020 unknown cause of death.This file will capture the first onset of biliary obstructions.Patient outcome: all three reinterventions were noted to be successful.On (b)(6) 2020, the patient died.The cause of death was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the evo-10-11-6-b device of lot number: c1237074 involved in this complaint was not returned for evaluation.With the information provided, a document-based investigation was conducted.This file is related to (b)(4) and was created in response to the (b)(4) study.Manufacturing records review: prior to distribution all devices are subjected to a visual inspection and functional inspection to ensure device integrity.A review of the manufacturing records did not reveal any discrepancies that could have contributed to this complaint issue.Instructions for use and/label review: it should be noted that the instructions for use (ifu0056) lists ¿stent occlusion¿ as a potential adverse event.There is no evidence to suggest that the customer did not follow the instructions for use.Image review: an image was not returned for evaluation.Root cause analysis: a definitive root cause could not be determined.A possible root cause can be attributed to patient pre-existing conditions and/or a known potential adverse event.From the study the cause of the obstruction was due to tissue or tumour ingrowth.Also, as previously noted, ¿stent occlusion¿ is listed as a potential adverse event in the ifu.Confirmation of complaint: the complaint is confirmed based on customer and/or rep testimony.Confirmed quantity of 01 x used device.Corrective action/correction: complaints of this nature will continue to be monitored for similar events.Summary of investigation: according to the initial reporter, the stent was implanted on (b)(6) 2016.The patient experienced abdominal pain and vomiting due to recurrent biliary obstruction 294 days post procedure.The patient required antibiotics and an endoscopic intervention to clean the prosthesis as a result of this occurrence.It was reported that the patient recovered/stabilised following this.Patient death was reported on (b)(6) 2020, the cause of death was unknown, from medical advisor input the death was related to stenosis of cancer.
|
| |
|
Event or Problem Description
|
|
A supplemental report is being submitted due to the completion of the investigation on 06-dec-2024.
|
| |
|