COOK INC RETRACTA DETACHABLE EMBOLIZATION COIL; KRD DEVICE, EMBOLIZATION, VASCULAR Back to Search Results
Model Number G34200
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/14/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
E1: customer (person): (b)(6).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned or that a death or serious injury occurred; nor is it admission that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
Cook was notified that during an embolization procedure where a retracta detachable embolization coil was being used, part of the wire guide with the coil attached separated.The clinician had difficulty with deployment of the device, there was an issue with releasing the retracta detachable embolization coil.The delivery wire was rotated counterclockwise many times over several minutes with the junction zone just inside the catheter tip.The device broke leaving part of the delivery wire and coil in the patient's vein.There is no plan to remove the portion of the delivery wire/coil from the patient.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Correction: d9, h3 and h6- annex a & annex g.Investigation: evaluation.Cook was notified that during an embolization procedure where a retracta detachable embolization coil was being used, part of the wire guide with the coil attached separated.The clinician had difficulty with deployment of the device, there was an issue with releasing the retracta detachable embolization coil.The delivery wire was rotated counterclockwise many times over several minutes with the junction zone just inside the catheter tip.The device broke leaving part of the delivery wire and coil in the patient's vein.There is no plan to remove the portion of the delivery wire/coil from the patient.Reviews of the documentation, including the complaint history, device history record, quality control procedures, and instructions for use (ifu), as well as visual inspection of the returned product, were conducted during the investigation.Cook received one delivery wire in a used and damaged condition.Upon visual inspection, cook confirmed that the delivery wire separated, and the junction zone was no longer attached to the wire.Additionally, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the device history record (dhr) for the device found no relevant nonconformance's that could have contributed to the reported failure mode.It should be noted that there were no other complaints associated with the final product lot number.Cook also reviewed product labeling: the current instructions for use [t_mwcer_rev6] state the following: "under fluoroscopic visualization, slowly advance the delivery wire until the entire length of coil exits the distal end of the catheter.Ensure that the junction remains positioned just inside the catheter tip.Note: advancing the delivery wire slowly allows the junction to be seen more easily and reduces the risk of damaging it.Note: if significant resistance is encountered during coil advancement, do not continue advancing.Retract the delivery wire slightly, then gently re-advance it.If there is still significant resistance, withdraw the delivery wire from the catheter and try using a new coil with a shorter length." evidence gathered upon review of the dmr, ifu, dhr, and returned device suggests that the device was not manufactured out of specification, and that there are no nonconforming devices in house or out in the field.Based on the information provided, examination of the returned product, and the results of our investigation, it was concluded that a component failure unrelated to manufacturing or design deficiencies contributed to the reported event.Per the risk assessment no further action is required.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETRACTA DETACHABLE EMBOLIZATION COIL
Common Device Name
KRD DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key20786227
Report Number1820334-2024-01553
Device Sequence Number12451331
Product Code KRD
UDI-Device Identifier10827002342009
UDI-Public(01)10827002342009(17)280922(10)15670724
Combination Product (Y/N)N
PMA/510(K) Number
K151676
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG34200
Device Catalogue NumberMWCER-35-14-12
Device Lot Number15670724
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/19/2024
Supplement Date Received by Manufacturer01/14/2025
Initial Report FDA Received Date11/26/2024
Supplement Report FDA Received Date01/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/22/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-