|
MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3507190MC |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
11/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Reason for device explant and/or reoperation: the patient has not undergone explantation or reoperation at this time.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast surgery with implantation of a 190cc mentor memorygel breast implant prosthesis.Post-operatively, the patient was diagnosed with a ruptured right breast implant using ultrasound imaging.A consultation with a medical professional has been conducted; however, at the time of this report, mentor has received no information regarding explantation or an expected explantation date.Several follow-ups have been conducted, and no additional information has been received.If more information becomes available, mentor will submit a supplemental report accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
On march 28, 2025, mentor was notified that the patient underwent bilateral removal and replacement with 700cc (left-sided) and 190cc (right-sided) mentor memorygel breast implants on (b)(6) 2025.Date of explantation: (b)(6) 2025.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On november 13, 2025, the mentor analysis lab received the suspect medical device for evaluation.On december 08, 2025, mentor completed an evaluation on the returned device.Device evaluation: mentor conducted a visual inspection.Visual analysis of the returned device revealed that the breast implant was found to have a tear on the posterior view between the union of the shell and patch.As the authorization form for examination was not received, the product evaluation lab could not reach a conclusion regarding the specific nature of the ruptures.The evaluation was limited to non-destructive testing.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: rupture can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.The following things may cause implant to rupture: damage by surgical instruments; stressing the implant during implantation and weakening it; folding or wrinkling of the implant shell; excessive force to the chest (e.G.During closed capsulotomy); trauma; compression during mammographic imaging; and severe capsular contracture.Breast implants may also simply wear out over time.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|