| Model Number |
816570 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The field service representative (fsr) reported that during preventative maintenance (pm) of the device, the roller pump belt was slipping.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The fsr replaced the belt and performed functional testing.The unit operated to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
| |
|