• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V+PRO, ENGLISH
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/30/2023
Type of Reportable Event Malfunction
Event or Problem Description
An authorized third party service provider (asp) contacted ventec to report that the device's exhaled tidal volume (vte) levels were low.There were no reports of patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
H6: the device was not returned to ventec for evaluation.The device was evaluated by an authorized service provider (asp) where the reported issue of its exhaled tidal volume (vte) levels being low was confirmed.The asp replaced the external flow module (efm) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the cause of the reported issue was the efm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
2603 s goyer rd
kokomo IN 46902
Manufacturer Contact
elizabeth gilbert
22002 26th ave se
bothell, WA 98021
4256861765
MDR Report Key20792636
Report Number3013095415-2024-00947
Device Sequence Number7809537
Product Code CBK
UDI-Device Identifier00850018761161
UDI-Public(01)00850018761161(11)200420
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K162877
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 11/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberV+PRO, ENGLISH
Device Catalogue NumberPRT-01185-001
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2024
Initial Report FDA Received Date11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/20/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-