BIOMET 3I OSSEOTITE¿ TAPERED CERTAIN¿ IMPLANT 5 X 8.5MM; DENTAL IMPLANT Back to Search Results
Catalog Number XIFNT585
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Osteolysis (2377)
Date of Event 09/05/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A4: patient weight: unknown / not provided.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
 
Event or Problem Description
It was reported a periimplantitis at tooth site #47.Implant was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE¿ TAPERED CERTAIN¿ IMPLANT 5 X 8.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key20794279
Report Number0001038806-2024-02755
Device Sequence Number7808578
Product Code DZE
UDI-Device Identifier00844868022046
UDI-Public(01)00844868022046(17)270506(10)2022031044
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K130949
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 11/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberXIFNT585
Device Lot Number2022031044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date11/27/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/06/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-