C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Catalog Number |
1808561 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Migration (4003)
|
| Health Effect - Clinical Codes |
Embolism/Embolus (4438); Thrombosis/Thrombus (4440)
|
| Date of Event |
12/14/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through litigation process that sometime later port placement procedure, the catheter fractured and the patient allegedly developed thrombosis.However, the current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation, medical report was provided for review.Medical record alleges that, the patient underwent an implantable port placement procedure using a powerport for chemotherapy delivery as part of lymphoma treatment.Eight months and seven days later, the patient underwent an xr chest, and it was observed that the fractured port a catheter with the distal component of the fracture had embolized into the left lower lobe pulmonary artery with possible associated thrombus.On the same day, it was further reported through imaging that technically successful main and left pulmonary angiography demonstrated a fracture catheter fragment in the left pulmonary artery.There was no significant pulmonary thrombus.The retrieval of the foreign body/catheter fragment from the left pulmonary artery was performed.Subsequently, the patient was planned for port removal.Therefore, the investigation is confirmed for the reported catheter fracture, migration and material separation.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B3, b5, g3, h6 (patient, device, method, result) section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported through litigation process that, on (b)(6) , 2021, a patient underwent port implantation via the right subclavian vein for the treatment of lymphoma.Approximately eight months and seven days later, on (b)(6) , 2021, the catheter had allegedly fractured with the distal component migrated and embolized to the left lower lobe pulmonary artery, which was confirmed by a chest x ray.Additionally, it was reported that the patient was diagnosed with thrombosis.Subsequently, the patient underwent port and fragment removal.However, the current status of the patient was unknown.
|
| |
|