C.R. BARD, INC. (BASD) -3006260740 PowerPort Implantable Port, Silicone Single-Lumen, 9.6F; Port & Catheter, Implanted, Subcutaneous, Intravascular
|
Back to Search Results |
|
| Catalog Number |
1759601 |
| Medical Device Problem Codes |
Fracture (1260); Difficult to Remove (1528); Material Separation (1562); Migration (4003)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
02/20/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported through litigation process that approximately one year, nine months and seven days later port placement procedure, the catheter allegedly fractured and the catheter fragment migrated to lung.The catheter fragment was unable to be retrieved because it had travelled to patient's pulmonary artery.The patient was then transferred to interventional radiology, where multiple unsuccessful attempts were made to retrieve the catheter fragment through groin.One day later, additional procedures were carried and the catheter fragment was successfully removed from the patient's pulmonary artery via groin.However, the current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|
|
|