| Model Number |
610-0211511-01 |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer's report was observed during initial testing; however, after disassembling the device to determine root cause, the report was no longer seen.The device was put through extensive testing including bench handling without duplicating the report.The processor/bridge/pace board was replaced as a precaution.The device passed the final test procedure, will be recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed a "defib disabled" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|