SYNTHES GMBH AIMING BLOCK F/SCR Ø5 F/LCP PAED-HIPPL; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.108.002
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/16/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported on november 16, 2024 device does not pass the kirschner needle 2.8 c/spatula tip through one of the holes of the guide block p/torn 5 p/ pl hip lcp pe.The surgeon had to modify his surgical planning and place a pediatric plaque of 3.5.
 
Event or Problem Description
Additional information received states that there was a 40 minutes surgical delay and the patient was in good condition.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Corrected: g1 and h6 (health effect - impact code and health effect - clinical code).H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h4, h6: the product was not returned to medtech orthopaedics, however photos were provided for review.The photo investigation of " 03.108.002, aiming block f/scr ø5 f/lcp paed-hippl" can not revealed the reported condition.As, the provided evidence was not sufficient to confirm the reported event of unable to assemble or disassemble.Functionality issues can not be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the observed condition of the aiming block f/scr ø5 f/lcp paed-hippl would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: product code: : 03.108.002; lot number : 2632984; release to warehouse date : 17.Aug.2010.Supplier: (b)(4).Manufacturing site: werk selzach.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9 h3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1.H3.H4, h6.The product was returned to j&j medtech orthopaedics for evaluation.The j&j medtech orthopaedics team conducted a visual inspection of the returned device.Visual inspection of the returned device found front hole of the aiming block is with signs of grooves/deformation making it visibly bigger leaving marks on the top and bottom of the block, surface of the block is lightly scratched as well causing to not assemble correctly into the mating device.The observed condition of the device is consistent with a component failure that was caused by exposure to unintended forces like overtightening the device while assemblance; or by assembling the device in a misaligned position.Properly handling and attention to the approved use of the device diminishes the risk of failure.A dimensional inspection was not performed due to it is not applicable to the complaint condition.A functional test was performed unsuccessfully due to the bent condition of the mating device.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the aiming block f/scr ø5 f/lcp paed-hippl would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: part# 03.108.002, lot # 2632984, manufacturing site: werk selzach logistik, supplier: (b)(4), release to warehouse date: 17 aug 2010, expiration date: na, a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: the product was not returned to medtech orthopaedics, however photos were provided for review.The photo investigation of " 03.108.002, aiming block f/scr ø5 f/lcp paed-hippl" can not revealed the reported condition.As, the provided evidence was not sufficient to confirm the reported event of unable to assemble or disassemble.Functionality issues can not be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the observed condition of the aiming block f/scr ø5 f/lcp paed-hippl would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review
=
> product code: (b)(4).Lot number: 2632984 release to warehouse date: 17.Aug.2010 expiration date: na supplier: synthes bettlach gmbh.Manufacturing site: werk selzach.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIMING BLOCK F/SCR Ø5 F/LCP PAED-HIPPL
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach CO 2425
SZ   2425
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key20795485
Report Number8030965-2024-14432
Device Sequence Number18090559
Product Code FZX
UDI-Device Identifier07611819847224
UDI-Public(01)07611819847224(10)2632984
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.108.002
Device Lot Number2632984
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/16/2024
Supplement Date Received by Manufacturer12/05/2024
05/26/2025
07/10/2025
09/25/2025
Initial Report FDA Received Date11/27/2024
Supplement Report FDA Received Date12/11/2024
05/27/2025
07/14/2025
09/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/17/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
K-WIRE Ø2.8 W/SPADE POINT TIP
Patient Age11 YR
Patient SexMale
-
-