MEDTRONIC, INC. MEDTRONIC CARELINK PROGRAMMER WITH ANALYZER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090AA4
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the programmer displayed autonomous cursor movement and the cursor jumped around the screen.It was noted that the programmer had printer issues.There was no patient involvement.
 
Additional Manufacturer Narrative
Product analysis: analysis was unable to confirm the customer comment that autonomous movement of the programmer cursor was observed.However, at analysis it was determined that the programmer failed the touchscreen test for unexpected/poor response to finger touch during operator interaction with touchscreen.Analysis was unable to confirm the customer comment that the programmer had printer issues.Electromagnetic interference was performed to resolve the touchscreen issue.Reconfigured hard drive, reloaded and updated software and the programmer then passed all final functional and systems tests.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC CARELINK PROGRAMMER WITH ANALYZER
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key20795864
Report Number2182208-2024-05032
Device Sequence Number7883002
Product Code KRG
UDI-Device Identifier00721902824692
UDI-Public00721902824692
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2090AA4
Device Catalogue Number2090AA4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/05/2024
Supplement Date Received by Manufacturer12/06/2024
01/10/2025
Initial Report FDA Received Date11/27/2024
Supplement Report FDA Received Date12/19/2024
01/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/29/2004
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2182208-071724-002-R
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2067L RADIO FREQUENCY HEAD, 229047 ANALYSER
Patient SexUnknown
-
-