• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON BD MINIDRAW¿ FINGER SLEEVE, SMALL; CAPILLARY BLOOD COLLECTION FUNNEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON BD MINIDRAW¿ FINGER SLEEVE, SMALL; CAPILLARY BLOOD COLLECTION FUNNEL Back to Search Results
Catalog Number 366613
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The manufacturing location for this product is roechling medical rochester (rmr).This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.There were multiple medical device types reported to be involved.The information for the additional device type is as follows: jka.There were multiple 510k numbers reported to be involved.The information for the additional 510k is as follows: k230493.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event or Problem Description
It was reported when using the bd minidraw¿ finger sleeve, small, there was sample (blood) pooling within an unspecified number of devices.There were no health consequences or impacts reported.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, 32 retention samples from bd inventory were evaluated by functional testing and 460 by visual inspection.No issues were observed as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode sample leakage.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported when using the bd minidraw¿ finger sleeve, small, there was sample (blood) pooling within an unspecified number of devices.There were no health consequences or impacts reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MINIDRAW¿ FINGER SLEEVE, SMALL
Common Device Name
CAPILLARY BLOOD COLLECTION FUNNEL
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
matt j. connor
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key20795876
Report Number2243072-2024-01172
Device Sequence Number7894429
Product Code GIM
UDI-Device Identifier50382903666138
UDI-Public(01)50382903666138(11)240206(17)241123(10)3143002
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K230391
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/23/2024
Device Catalogue Number366613
Device Lot Number3143002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/12/2024
Supplement Date Received by Manufacturer08/15/2025
Initial Report FDA Received Date11/27/2024
Supplement Report FDA Received Date08/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/06/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-