The eli 380 is intended to be a high-performance, multichannel resting electrocardiograph.As a resting electrocardiograph, the eli 380 simultaneously acquires data from each lead.Once the data is acquired, it can be analyzed, reviewed, stored, printed, or transmitted.It is a device primarily intended for use in hospitals but may be used in medical clinics and offices of any size.Device is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.Investigation of the device found issue with software.Although the reported event did not result in a serious injury or death, the report of a patient demographics mismatch in ecg tests could lead to a misdiagnosis or mistreatment, if it were to recur.Therefore, baxter considers this complaint reportable.
|