| Model Number |
176645 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/20/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant medical products: 176645, 176645 endo dissect 254 (lot# p4g1143), 176645, 176645 endo dissect 254 (lot# p4g1143) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a right hemicolectomy, a laparoscopic procedure, the jaws of the three hand instruments did not fit or close properly.The issue was detected at the start of the surgery when the device was removed from the packaging.An inventoriable device was used to complete the surgery.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.A visual inspection of the returned picture noted the image depicts the jaws of the device closed with a gap at the proximal end.Visual inspection of the product noted that the jaws were misaligned.The instrument was applied to test media.The jaws opened and closed without difficulty.It was reported that the jaws will not close completely.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the device is exposed to a side force (leverage) that consequently breaks one side of the jaws.Another possibility is when the device is activated with too much force to the jaws and is used in a twisting motion resulting in breakage.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.Consult the medical literature relative to complications, hazards and techniques, prior to performing endoscopic procedures.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|