| Catalog Number |
305194 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Material #:305194 , batch#:3347792.Verbatim: issue: clogged/blocked needle.No pt harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial mdr submission.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
No additional information received.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Follow up for device evaluation.It was reported the needle was clogged.To aid in the investigation, one hundred thirty-one samples in sealed packaging blisters were received for evaluation by our quality team.Eighty samples were randomly selected for investigation.A visual inspection was performed, and no defects or imperfections were observed.Each sample was then assembled to a syringe with saline solution.All the samples expelled the solution with a normal flow.A device history record review was completed for provided material number 305194, lot 3347792.The review did not reveal any detected quality issues during the production of this lot that could have contributed to the reported defect.There were no related quality notifications.All processes and final inspections complied with specification requirements.Based on the investigation and with the returned sample analysis the symptom reported by the customer could not be confirmed.
|
| |
|