Model Number 11500A
Medical Device Problem Code Device Stenosis (4066)
Health Effect - Clinical Codes Dyspnea (1816); Fatigue (1849)
Date of Event 11/04/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 19mm 11500a aortic valve in aortic position was disabled via valve in valve procedure after an implant duration of 4 years, 7 months due to aortic stenosis.The patient presented with shortness of breath and fatigue.Tavr was performed with a 20mm 9750tfx transcatheter valve and the patient was stable at the end of the procedure.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the subject device is not available for evaluation, as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated sections: b4, g3, g6, h2, h6 (type of investigation, investigation findings, and investigation conclusions).The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Engineering evaluation summary: per (b)(4), stenosis is a common manifestation of bioprosthetic valve and valved conduit dysfunction and has many potential root causes, including structural valve deterioration (svd), non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis.Bioprosthetic device stenosis is typically related to patient and/or procedural factors and is not typically an indication of a device malfunction related to a manufacturing deficiency.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient-related factors, including hyperlipidemia and coronary artery disease.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INSPIRIS RESILIA AORTIC VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2- office m2013
irvine, CA 92614
9492506615
MDR Report Key20799321
Report Number2015691-2024-09031
Device Sequence Number17824232
Product Code LWR
UDI-Device Identifier00690103194968
UDI-Public(01)00690103194968(17)210129
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P150048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/29/2021
Device Model Number11500A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/04/2024
Supplement Date Received by Manufacturer12/11/2024
Initial Report FDA Received Date11/27/2024
Supplement Report FDA Received Date01/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/30/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age83 YR
Patient SexMale
-
-