• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3505350BC
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: n/a.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 350cc mentor memorygel breast implant ruptured intra-operatively, during a primary breast augmentation.There was no report of adverse event or patient consequence.Another unspecified implant was placed.
 
Additional Manufacturer Narrative
On january 28, 2025, the mentor failure analysis lab received the device for evaluation.On february 4, 2025, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and microscopic examination of the returned device.Visual analysis of the returned device determined that the smooth uhp 350cc breast implant had an area of missing material on the posterior view, measuring approximately 1.0 cm x 1.0 cm.Microscopic examination was performed on the edges of the missing material, and parallel striations were found in an area of the missing material, measuring approximately 0.1 cm.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.The cause of the rupture in the remaining area of the tear could not be identified.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key20799559
Report Number1645337-2024-13984
Device Sequence Number20580348
Product Code FTR
UDI-Device Identifier00081317006923
UDI-Public(01)00081317006923(17)280515(10)9906930
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3505350BC
Device Lot Number9906930
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/06/2024
Supplement Date Received by Manufacturer01/28/2025
Initial Report FDA Received Date11/27/2024
Supplement Report FDA Received Date02/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-