| Model Number |
FORCEFX-8C |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during use, the electrocoagulation wire connected to the electrocoagulation hook caught fire with a "pop" sound, and soon the electrocoagulation wire was burned off.The operator hurried to put out the fire.
|
| |
|
Search Alerts/Recalls
|