| Model Number |
030459 |
| Medical Device Problem Codes |
Failure to Form Staple (2579); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
11/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during a thoracoscopic segmentectomy, when removal of lung tissue, the stapler was able to fire, the staples were not completely stapled on the lung, and all of them fell off, and the staple line was incomplete centrally.It was noted that no component fell into the cavity of the patient.Another reload was used, but same issue occurred.A replacement device from another manufacturer was done to resolve the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: egiaustnd, egiaustnd endogia ultra univ std stap (lot#:p4d1113s), 030459, 030459 endo gia r/or 60 4.8mm (lot#:t1j195x).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a thoracoscopic segmentectomy, when removal of lung tissue, the stapler was able to fire, the staples were not completely stapled on the lung, and all of them fell off, and the staple line was incomplete centrally.Another reload was used, but same issue occurred.A replacement device from another manufacturer was done to resolve the issue. .
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the loading unit was partially fired to the 5cm cut line.The clamping mechanism was deformed.It was reported that that the staple line was incomplete and a component disengaged.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if: application over tissue that was beyond the recommended thickness range.Application with an obstacle incorporated in the jaws.In any of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ens ure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: select a loading unit with the appropriate staple size for the tissue thickness.Overly thick or thin tissue may result in unacceptable staple formation.Always include the combined thickness of the tissue and of any staple line reinforcement material in use when choosing the proper staple cartridge.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|