| Model Number |
030457 |
| Medical Device Problem Codes |
Failure to Form Staple (2579); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
11/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic radical resection of esophagogastric junction cancer, a stapler was used when the duodenum was detached. the suture line was incomplete, the staples did not form into a good b-shape, bleeding occurred, and it was finally sutured manually. during the r-y anastomosis of the esophagus and jejunum, the eea21 could not be fired normally.The operation was completed after replacing it with a new handle.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: eea21, eea21 eea 21mm sgl use stapler, (lot #p0c1630ky); egiaustnd, egiaustnd endogia ultra univ std stap, (lot #p2c0582s).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic radical resection of esophagogastric junction cancer, a stapler was used when the duodenum was detached.The suture line was incomplete, the staples did not form into a good b-shape, bleeding occurred, and it was finally sutured manually.During the r-y anastomosis of the esophagus and jejunum, the circular stapler could not be fired normally.The operation was completed after replacing it with a new handle.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5, g3, h6 (ed, imf) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the loading unit was partially fired to the 2cm cut line.The anvil was bowed.The device had a buckled knife bar assembly.It was reported that the suture line was incomplete, the staples did not form into a good b-shape.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if application over tissue that is beyond the recommended thickness range.Application with an obstacle incorporated in the jaws.In any of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly and tissue may not be fully transected.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: select a loading unit with the appropriate staple size for the tissue thickness.Overly thick or thin tissue may result in unacceptable staple formation.Always include the combined thickness of the tissue and of any staple line reinforcement material in use when choosing the proper staple cartridge.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|