• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; BREATHING CIRCUIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER & PAYKEL HEALTHCARE; BREATHING CIRCUIT Back to Search Results
Model Number RT266
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/01/2024
Type of Reportable Event Malfunction
Event or Problem Description
A distributor reported on behalf of a healthcare facility in japan that an rt266 infant dual-heated evaqua2 breathing circuit failed the ventilator leak test prior to patient use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).The subject rt266 breathing circuit is currently en route to nz for evaluation.We will provide a follow up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(4).Method: the subject rt266 breathing circuit was received at fisher & paykel (f&p) healthcare in new zealand for evaluation, where it was visually inspected and analysed.Results: visual inspection of the tubing did not identify any visible deformities; however, leak testing confirmed the presence of leaks from micro voids in the tubing.Conclusion: the reported leak was confirmed.All rt266 breathing circuits are visually inspected, and pressure and flow tested during production and those that fail are rejected.The subject circuit would have met the required specifications at the time of production.The user instructions that accompany the rt266 breathing circuit state: perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient.Visually inspect breathing sets for damage (e.G.A crushed tube or cracked connector) before use and replace if damaged.
 
Event or Problem Description
A distributor reported on behalf of a healthcare facility in japan that an rt266 infant dual-heated evaqua2 breathing circuit failed the ventilator leak test prior to patient use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER & PAYKEL HEALTHCARE
Common Device Name
BREATHING CIRCUIT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key20803079
Report Number9611451-2024-00878
Device Sequence Number9934980
Product Code BZE
UDI-Device Identifier09420012431103
UDI-Public(01)09420012431103(10)2103089408(11)240416(17)290416
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103767
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot Number2103089408
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/01/2024
Supplement Date Received by Manufacturer02/04/2025
Initial Report FDA Received Date11/27/2024
Supplement Report FDA Received Date03/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-